Images for Press Release

Agilent Technologies' New Source-Corrected Noise-Figure Measurement Option Extends Single-Connection Multiple-Measurement Capabilities of the PNA-X Network Analyzer

The following product photos are available for use by the media, based on the Keysight Photography Use Policy.

With a single connection to the amplifier, the PNA-X network analyzer offers fast, high accuracy measurements of S-parameters, noise figure, compression, IMD, harmonics, and more.

Image 1:
With a single connection to the amplifier, the PNA-X network analyzer offers fast, high accuracy measurements of S-parameters, noise figure, compression, IMD, harmonics, and more.

Download JPEG (0.7MB)

Download TIF (10.3MB)

Agilent's  unique noise-figure-calibration technique uses an ECal module as an impedance tuner to remove the effects of imperfect system source match. The result is noise-figure-measurement accuracy that surpasses standalone Y-factor-based noise figure analyzers or spectrum analyzer solutions.

Image 2:
Agilent's unique noise-figure-calibration technique uses an ECal module as an impedance tuner to remove the effects of imperfect system source match. The result is noise-figure-measurement accuracy that surpasses standalone Y-factor-based noise figure analyzers or spectrum analyzer solutions.

Download JPEG (4.2MB)

Download TIF (10.4MB)

An example measurement of fully error-corrected amplifier gain (S21) and noise figure from 500 MHz to 26.5 GHz, performed with a single connection to the amplifier.

Image 3:
An example measurement of fully error-corrected amplifier gain (S21) and noise figure from 500 MHz to 26.5 GHz, performed with a single connection to the amplifier.

Download PNG (0.1MB)

Download TIF (0.1MB)

The PNA-X is the new standard in microwave network analysis.

Image 4:
The PNA-X is the new standard in microwave network analysis.

Download JPEG (1.2MB)

Download TIF (5.9MB)

An example screen image of the PNA-X measuring S-parameters (upper left), gain and phase compression (upper right), intermodulation distortion (lower left), and noise figure (lower right), all with a single connection to the amplifier under test.

Image 5:
An example screen image of the PNA-X measuring S-parameters (upper left), gain and phase compression (upper right), intermodulation distortion (lower left), and noise figure (lower right), all with a single connection to the amplifier under test.

Download PNG (0.7MB)

Download JPG (0.5MB)

Agilent's  unique noise-figure-calibration technique uses an ECal module as an impedance tuner to remove the effects of imperfect system source match, thereby providing accurate measurements of 50-ohm noise figure. Measurement accuracy surpasses standalone Y-factor-based noise figure analyzers or spectrum-analyzer-based solutions.

Image 6:
Agilent's unique noise-figure-calibration technique uses an ECal module as an impedance tuner to remove the effects of imperfect system source match, thereby providing accurate measurements of 50-ohm noise figure. Measurement accuracy surpasses standalone Y-factor-based noise figure analyzers or spectrum-analyzer-based solutions.

Download JPEG (0.3MB)

Download TIF (4.6MB)

PNA-X source-corrected noise figure measurements expand Agilent's leadership and excellence in noise-figure analysis.

Image 7:
PNA-X source-corrected noise figure measurements expand Agilent's leadership and excellence in noise-figure analysis.

Download JPEG (0.3MB)

Download TIF (3.0MB)

With a single connection to the amplifier and a unique noise-figure-calibration technique that uses an ECal module as an impedance tuner, the PNA-X offers high accuracy noise figure measurements as well as measurements of S-parameters, compression, harmonics, and IMD.

Image 8:
With a single connection to the amplifier and a unique noise-figure-calibration technique that uses an ECal module as an impedance tuner, the PNA-X offers high accuracy noise figure measurements as well as measurements of S-parameters, compression, harmonics, and IMD.

Download JPEG (0.1MB)

Download TIF (0.9MB)

 

Related Information
  Press Release:

Agilent Technologies Extends High-Performance Noise-Figure Measurement Technique to 50 GHz
(2012-November-28)

Agilent Technologies' New Source-Corrected Noise-Figure Measurement Option Extends Single-Connection Multiple-Measurement Capabilities of the PNA-X Network Analyzer
(2007-September-03)

  Keysight Photography Use Policy:

We will permit use of Keysight photography available via this site as long as the use will not disparage Keysight nor imply endorsement by Keysight. The photography cannot be changed or manipulated in any way with the exception of resizing or cropping. By downloading our product photography you are indicating that you accept these terms. Any commercial use of these images must be approved by Keysight Technologies.

ALL IMAGES USED MUST INDICATE:
"Reproduced with Permission, Courtesy of Keysight Technologies, Inc."

  How do I download images:

To download files in Windows, use the right mouse button on the desired format (JPEG or TIFF) and choose 'Save Target As...'